45 Degree Flammability Tester maintenance how to do:


1, the combustion tester gas tank should have a pressure reducing valve, the main valve off start reliable.


2, combustion tester combustion test should be operated in the presence of two people, one is responsible for the operation of the sample test, the other is responsible for supervision, check whether the gas pipeline before the test leakage, if necessary, open the pneumatic small box to check, such as joints, gas pipe is sealed to avoid accidents, the test, the tester responsible for supervision zui close to the main valve of the gas tank, found abnormalities in time to close the main valve to find the cause, the test After the test is over, be sure to close the main valve.


3, combustion tester test chamber should be installed in the air convection but can be naturally ventilated test cabinet, the test to prevent air flow, to avoid the burner backfire. The end of the test can start the fan to remove the smoke.

Horizontal Burning Tester.jpg

4, combustion tester input power supply for AC220V ± 10%, the application of grounded three-pole socket, to ensure reliable grounding.


5, the combustion tester electric control, must be time to reach the time relay control burner back to the original position.


6, combustion tester test room should be equipped with fire extinguishing equipment. Regularly check the gas circuit.


7, should not make the fire nozzle by the impact of heavy objects or vigorous bending, so as not to damage the fire nozzle affect the test.


8, the use of heavy hammer stop switch, the hammer does not need to be suspended too high, in order to protect the stop switch.